TURKISH JOURNAL OF ONCOLOGY 2012 , Vol 27 , Num 4
Evaluation of the dose distribution behind the prostheses in prostate cancer patients with hip prostheses using film dosimetry and specially designed phantom
Emre YURTÇU, Sinan HOCA, Nezahat OLACAK, İbrahim OLACAK, Yusuf Ziya HAZERAL, Arif Bülent ARAS
Ege Üniversitesi Tıp Fakültesi Hastanesi, Radyasyon Onkolojisi Kliniği, İzmir OBJECTIVES
We aimed to investigate the effect of hip prostheses on dose distribution in radiotherapy patients.

METHODS
Solid Water phantom-containing prostheses (cobalt-chromium- molybdenum and cobalt-chromium-molybdenum-titanium) were placed between the phantoms used for the measurements. Films at 10, 20 and 30 cm depths were irradiated for 6 and 18 MV with 50 cGy dose in the presence or not of prosthesis. Dose profiles were evaluated statistically.

RESULTS
For 6 and 18 MV photon energies, significant differences was found along the thick and vertical axes of both prostheses at 10 and 20 cm depths and along the middle, thick and vertical axis at 30 cm depth (p<0.05).

CONCLUSION
Hip prostheses affect radiotherapy dose distribution, and this should be considered when planning a treatment. Keywords : Dose profiles; hip prosthesis; radiotherapy