Dosimetric and Mechanical Stability of CyberKnife Robotic Radiosurgery Unit: 5 Years' Clinical Experience
Fatih BİLTEKİN, Gözde YAZICI, Mustafa CENGİZ, Ali DOĞAN, Bülent ÜNLÜ, Mete YEĞİNER, Gökhan ÖZYİĞİT
Department of Radiation Oncology, Hacettepe University Faculty of Medicine, Ankara-Turkey OBJECTIVE
The purpose of the present study was to investigate dosimetric and mechanical stability of CyberKnife Robotic Radiosurgery System (Accuray Inc., Sunnyvale, CA, USA) in short- and longterm period.

METHODS
Output factor measurements and automated quality assurance (AQA) tests performed on CyberKnife unit 2009-2013 at radiation oncology department of Hacettepe University, Turkey, were analyzed retrospectively.

RESULTS
According to the analysis, more than 95% of the output measurements over 5 years were within the tolerance limit ≤2%. In AQA test analysis, 144 AQA test results were within the tolerance limit from 2009 to 2011. However, 7 of the 51 measurements taken in 2012, and 4 of the 47 measurements performed in 2013 exceeded 1 mm radial error.

CONCLUSION
Output and AQA data of CyberKnife system indicate that it is quite stable in daily and long-term period. Nevertheless, daily measurements should be performed on CyberKnife unit since high radiation dose per fraction is usually delivered to target volume. Keywords : CyberKnife unit; output stability; stereotactic radiosurgery