The purpose of the present study was to investigate dosimetric and mechanical stability of CyberKnife Robotic Radiosurgery System (Accuray Inc., Sunnyvale, CA, USA) in short- and longterm period.
METHODS
Output factor measurements and automated quality assurance (AQA) tests performed on CyberKnife
unit 2009-2013 at radiation oncology department of Hacettepe University, Turkey, were
analyzed retrospectively.
RESULTS
According to the analysis, more than 95% of the output measurements over 5 years were within the
tolerance limit ≤2%. In AQA test analysis, 144 AQA test results were within the tolerance limit from
2009 to 2011. However, 7 of the 51 measurements taken in 2012, and 4 of the 47 measurements
performed in 2013 exceeded 1 mm radial error.
CONCLUSION
Output and AQA data of CyberKnife system indicate that it is quite stable in daily and long-term
period. Nevertheless, daily measurements should be performed on CyberKnife unit since high radiation
dose per fraction is usually delivered to target volume.