TURKISH JOURNAL OF ONCOLOGY 2015 , Vol 30 , Num 2
Evaluation of cylindrical diode array performance for VMAT (Volumetric Arc Therapy) verification
Hande BAŞ AYATA, Cemile CEYLAN, Ayhan KILIÇ, Timur UĞUR, Metin GÜDEN, Kayıhan ENGİN
Özel Anadolu Sağlık Merkezi, Radyasyon Onkolojisi Kliniği, Kocaeli OBJECTIVES
In our study, we were performed tests to understand the characteristics of the ArcCHECK QA system and were evaluated the suitability of this QA system for VMAT verification.

METHODS
Short term reproducibility, dose linearity, dose rate and field size dependence tests were performed to evaluate the characteristics of the ArcCHECK QA system.

RESULTS
For the short term reproducibility, the standard deviations were found 0.043% for 6X, 0.06% for 10X and 0.057% for 6FFF. For the dose linearity, when its output was greater than 20 MU, the dose response variations were within ±0.3% for 6X, ±0.2% for 10X and ±0.16% for 6FFF. For the dose rate dependence, the dose responses exhibited 0.6% for 6X, 1.1% for 10X and 0.4% for 6FFF with increasing the dose rate.

CONCLUSION
The short term reproducibility, the dose linearity, the dose rate and field size dependence of ArcCHECK are very good and it is suitable for pretreatment quality assurance of VMAT plans. Keywords : Diod array; quality assurance; IMRT; VMAT