TURKISH JOURNAL OF ONCOLOGY 2013 , Vol 28 , Num 2
Midplane dose determination and verification of calculated doses in total body irradiation
Özlem ÖZDEMİR,1 Sinan HOCA,2 Nezahat OLACAK,2 Serdar ÖZKÖK,2 Serra KAMER,2 Yavuz ANACAK2
1Pamukkale Üniversitesi Tıp Fakültesi, Radyasyon Onkolojisi Anabilim Dalı, Denizli
2Ege Üniversitesi Tıp Fakültesi, Radyasyon Onkolojisi Anabilim Dalı, İzmir
OBJECTIVES
To compare calculated and measured doses for different regions of anthropomorphic phantom and patients using ion chamber and thermoluminescence dosimetry (TLD) for total body irradiation.

METHODS
Measurements were done for lateral fields with 6 MV, gantry 82º, 40x40 cm2 field and 400 cm source-axis distance (SAD). Entrance-exit and midline doses were measured on anthropomorphic phantom by TLD and entrance-exit doses were measured by TLD and ion chamber on patients.

RESULTS
For anthropomorphic phantom measurements differences between calculated and measured entrance-exit doses of head, neck, shoulder, lung and thick pelvis were 0.8%, 2.7%, 26.4%, 4.4% and 4.9% and for midline doses were 1.6%, 1.6%, 6.3%, -1.4% and 7.4% respectively. For patients; TLD differences were within -4.13% ile 6.7%, -3.3% ile 3.9%, 5.1% ile 16.6%, -7.8% ile 2.4%, and 3.6% ile 7.1% respectively. For thick pelvis measurements with ion chamber differences were within %0.1-1.9.

CONCLUSION
Total body irradiation is being applied in limit values in our clinic. Keywords : Total body irradiation, in-vivo dosimetry, thermoluminescence dosimetry