The purpose of this study is providing dosimetric control by comparing dose values calculated in computerized treatment planning system (CTPS) and measured by Linear 2D-Array and ion chambers of various volumes.
METHODS
Fields measured by Farmer, Semiflex and PinPoint ion
chambers were created in CTPS. Using Linear 2D-Array,
measurements were done in open, isocentric, oblique and
wedge filtered fields which created in CTPS. Dose values were
compared.
RESULTS
PinPoint ion chamber measured top dose values whereas
Farmer ion chamber measured low values. The most compatible
results with CTPS were obtained by Semiflex ion chamber.
However, in 2D-Array measurements, concordance was
observed with CTPS in central axis, difference reached 20 %
in build-up and penumbra regions.
CONCLUSION
The most compatible results with CTPS were obtained by
Semiflex ion chamber. Reasons of discordance in 2D-Array
measurements were found as; varible dose around build-up
region, and not to obtain the exact superpose of profiles' penumbra
region.